Abstract

A multislice gradient echo sequence (FLASH) was compared with a conventional spin-echo (SE) technique with regard to its value for contrast enhanced MR diagnosis. In 28 patients with cerebral tumours, SE images (SE 400/30; four images/3.4 minutes) and FLASH images (FLASH 315/14; 15 images/1.4 minutes) were obtained before and after gadolinium DTPA. After gadolinium-DTPA results were comparable for both techniques with respect to contrast enhancement, tumor contrast and delineation. Because of the higher efficiency of the FLASH 315/14 technique, this sequence is the method of choice for contrast enhanced cerebral MR imaging.